| DFS-II Engineering Specifications |
Greyline DFS-II Doppler Flow Switch
Standard features include : |
 |
Sensor - strap-on, single head ultrasonic |
 |
Sensor cable - 20ft (6m) shielded coaxial pair |
 |
Installation - sensor compound kit and stainless steel pipe clamp |
 |
Enclosure - watertight NEMA4X (IP67) polycarbonate |
 |
1 relay - 5 amp DPDT |
 |
Set point - ON/OFF adjustment 0.25 to 10 ft/sec (0.076 to 3 m/sec) |
 |
Failsafe mode - switch selectable |
 |
Indication - relays and signal strength LED's |
 |
Sensitivity - adjustable |
 |
Time delay - adjustable, 0-60 seconds |
 | Power input - 120VAC 50/60Hz |
 |
Electrical surge protection
|
 |
1 manual - Installation and Operation
|
|
| |
OPEN CHANNEL FLOW MONITOR SPECIFICATIONS
|
SCOPE : This specification covers an ultrasonic, echo-ranging type open channel flow meter as manufactured by Greyline Instruments, Massena, New York / Long Sault, Ontario. This instrument shall provide for indicating, transmitting, totalizing and data logging of the flow rate through a flume, weir, or other primary measuring device.
|
A. GENERAL |
 |
Flow Monitor to consist of an non-contacting ultrasonic sensor, connecting cable, and a remote enclosure with indicating, transmitting, controlling and data logging electronics.
|
 |
Measurement accuracy shall be ±0.25% and shall be automatically temperature compensated.
|
 |
Sensor cable length shall be as required by installation, not to exceed 500 ft (152 m).
|
 |
System shall have no moving parts and shall not contact flowing water.
|
|
B. SENSING ELEMENT
|
 |
Sensor shall withstand accidental submersion to 20 psi.
|
 |
Sensor minimum range shall be 8" (203.2 mm) and maximum range shall be 15 ft (4.57 m).
|
 |
Sensor shall be constructed of PVC.
|
 |
Sensor operating temperature shall be from -40°F to 150°F (-40°C to 65°C).
|
 |
Sensor shall include integral temperature sensor. Temperature sensors requiring separate mounting and wire runs shall not be accepted.
|
|
C. SENSOR CONNECTING CABLE
|
 |
Provide RG62AU coaxial cable, minimum length 25 ft (7.6 m), with waterproof, potted bond to the Sensor head. |
 |
Level and temperature signals shall be conducted on one single coaxial cable. Separate or multiple-conductor cables shall not be accepted. |
 |
Extended sensor cable shall be RG62AU coaxial to a maximum of 500 ft (152m). Cable shall be spliced with screw terminal connections in manufacturers recommended steel NEMA4 Junction Box. |
 |
Sensor cable shall be installed in grounded metal conduit. |
|
D. TRANSMITTER
|
 |
The transmitter shall provide for field calibration to common primary metering devices, plus allow entry of "K" and "n" values for calibration to nonstandard flumes, weirs or open channels. |
 |
Field-calibration shall be via built-in 3-key calibration system with menu selection of parameters. Systems requiring calibration by Parameter codes, BCD switches or external calibrators shall not be accepted. |
 |
Calibration data shall be password protected and permanently stored through power interruptions for a minimum of 12 months. |
 |
Field calibration shall allow selection and automatic conversion of measurement units, measurement span, high/low flow alarm relay and flow proportional relay pulse rates. |
 |
Transmitter shall permit field programmable damping to smooth output in turbulent flow conditions, and programmable rejection time to disregard false signals from rain, wind, floating objects and spurious echoes. |
 |
Transmitter operating temperature shall be from -5° to 140°F (-20° to 60°C). Transmitter shall contain a thermostat-controlled enclosure heater for condensation protection below 30°F (-1°C). |
 |
Have a built-in 50,000 point data logger with USB and RS232C output and connecting cables. Data logger shall support time and date-stamped logging and display daily flow reports including total, average, minimum, maximum and times of occurrence. |
 |
Include Windows software for data log retrieval, graphing and export. |
 |
Transmitter shall have an isolated 4-20mA output into a 1000 ohm maximum load and user-selectable 0-5V output. |
 |
Provide 3 relay contacts rated 5 amp SPDT programmable to operate a remote totalizer, a sampler, high-low flow alarm plus echo loss alarm. |
 |
Provide a 4-digit LCD display scaled in user-selected engineering units and visible without opening the Transmitter enclosure. |
 |
Provide a 16-digit alphanumeric display for calibration menu, running totalizer, flow reports, or error message display of echo loss or other fault conditions. |
 |
Transmitter shall be housed in a wall-mount, watertight NEMA4X (IP66) enclosure with hinged, clear cover. Mounting hardware shall be included. |
 |
Transmitter Displays indicating flow rate, units of calibration, maximum flow rate and flow totalizer shall be visible without opening cover. |
 |
Transmitter electronics shall be surge protected on AC power input, sensor, 0-5V and 4-20 mA output. |
 |
Transmitter power input shall be 115VAC 50/60Hz with power consumption of 6.2 Watts or less.
|
|
E. OPTIONAL FEATURES FOR INSERTION AS REQUIRED
|
 |
Electronics enclosure shall include an Intrinsic Safety Barrier. Sensor, connecting cable and junction boxes shall be rated intrinsically safe to Class I,II,III, Div. I,II, Groups C,D,E,F,G.
|
 |
Have a thermostat controlled Enclosure Heater for transmitter operation at temperatures below freezing.
|
 |
Have manufacturer's recommended Enclosure Sunscreen to permit electronics enclosure mounting in direct sunlight.
|
 |
Have manufacturer's recommended Sensor Sunscreen to permit temperature-compensated sensor mounting in direct sunlight.
|
 |
Have manufacturer's recommended sensor mounting stand.
|
 |
Transmitter power input shall be 200-260VAC 50/60Hz with maximum power consumption of 6.2 Watts or less.
|
 |
Transmitter power input shall be 9-36VDC with minimum power consumption of 2.1 Watts.
|
|
F. MANUFACTURER
|
Open Channel Flow Monitor shall be Greyline Model OCF 4.0 and warranted against defects in materials and workmanship for one year.
|
|
| |
|
|
|